Sigma 6k15 Centrifuge



Centrifuge

  1. Sigma 6k15 Centrifuge Machine
  2. Sigma 6k15 Centrifuge Manual
  3. Sigma 6k15 Centrifuge Model
One moment please...
Not A Member?Sign Up
Sigma 6k15 Centrifuge0 Replies

View & download of more than 513 Sigma PDF user manuals, service manuals, operating guides. Camera Lens, Bicycle Accessories user manuals, operating guides & specifications. View & download of more than 513 Sigma PDF user manuals, service manuals, operating guides. Camera Lens, Bicycle Accessories user manuals, operating guides & specifications.

Do you have a estrutural plattform where is to mounten all devices of the OIL CENTRIFUGE SIGMA 6-15H model: '6-15/9 K15 - 09/05/01' ??

RE: Sartorius - SIGMA 6-15 and 6K151 Reply

SIGMA 6-15 CENTRIFUGE?

erro 25 !a drive problem ! Someone could help me to solve this problem ??

RE: Sartorius - SIGMA 6-15 and 6K151 Reply

GOSTARIA DE SABER QUAL É O TEMPO DE AQUECIEMENTO PARA ATINGIR 60C NESTA centrifuga MODELO 6-15h.

6k15

E QUAL A TEMPERATURA NORMAL DE TRABALHO DA MESMA.

GRATO,

UDENIR

RE: Sartorius - SIGMA 6-15 and 6K15
1 ReplyI HAVE A SIGMA 6K15 CENTRIFUGE, IT HAS A PROBLEM WITH THE DOOR. WHEN THE DOOR IS CLOSED THE SENSOR START AND CLOSE THE DOOR, IN A FEW SECONDS THE DISPLAY SENDS A MESSAGE 'PROBLEM WITH THE DOOR OPEN. TURN ON AND OF THE POWER SWITCH' PROBLEM 78. I FOLLOW THE INSTRUCTION THAT APPEARS IN THE DISPLAY AND THE PROBLEM STARTS AGAIN AND AGAIN. I REVIEWED THE MANUAL AND THIS PROBLEM DOES NOT APPEAR SO I NEED YOUR HELP FOR RESOLVE THIS PROBLEM. THANKS[i]Originally asked by HECTOR VALDEZ via email[/i] RE: Sartorius - SIGMA 6-15 and 6K15
Sigma 6k15 Centrifuge
0 RepliesPlease, My question isWaht to do with TH e error 43, imbalance offset fail? If we switch on the centrifuge 6K15 and before deposing any plate we get this error? Thank you for your help RE: Sartorius - SIGMA 6-15 and 6K15
Sigma 6k15 centrifuge definition

General Info

Sigma 6k15 Centrifuge Machine

This SOP is a how to use guide for the Sigma 4K15 Centrifuge.

Procedure

Sigma 6k15 Centrifuge Manual

  1. Turn on the centrifuge using the mains power switch located on rear back panel.
  2. If the centrifuge lid is closed then gently press the Lid-Key (Illuminated Orange), located on the top right hand side of the main control panel.
    1. Wait for the lid to unlock and lift fully open.
  3. Check that there are two black sample wells and two metal Plate holders located inside the machine.
  4. Gently remove the plate holders from the sample wells and place to one side.
  5. Taking care, position your plates in the metal plate holders so that they are correctly aligned with one another. You may load up to 3 plates in each holder.
    1. Never run the centrifuge with an unequal number of plates in each holder. ##Always balance out odd numbers with an empty plate rack.
    2. Failure to do so will cause the run to fail and cause damage to the centrifuge.
  6. Lower the loaded holders back into position ensuring that they are as far into the black sample wells as possible.
    1. Do not use excessive force to replace the holders. If the fit is not correct, check the alignment of your plates.
  7. Lower the centrifuge lid back into the closed position and hold until you hear it lock into place.
    1. Both the Lid-key (illuminated Orange) and the Start-key (illuminated green) should now be highlighted.
  8. Using the blue circular knob located on the front of the centrifuge display, highlight the ‘speed’ panel. Depress the knob once and the word SET will appear inverted. Adjust to the desired speed, measured as rpm, by turning to the left or right.
    1. Minimum value of 100 rpm, Maximum value of 6200 rpm.
  9. Press the blue circular knob again to set this value.
  10. Follow the same procedures as detailed in section 9 to set both the time and temperature which you wish to centrifuge at.
    1. Temperature range: -20 to 40°c
    2. Time range: 1 minute and upwards
  11. Alternatively, select the desired program number as above.
  12. Once the correct parameters have been stored for your run, press the Start-key (illuminated green). The centrifuge will now run through your requested method.
  13. Once the cycle has completed, open the lid of the centrifuge by pressing the Lid-key (illuminated orange).
    1. To stop the cycle at ANY time during the run, press the Stop-key (illuminated red)
  14. Carefully remove plates in holders from the centrifuge. Remove plates from plate holders and replace holders back into black sample wells.

Sigma 6k15 Centrifuge Model

Retrieved from 'https://openwetware.org/mediawiki/index.php?title=CRI_Sigma_4K15_centrifuge&oldid=80128'